CVE-2019-25719
High · 8.6CVSS
8.6
EPSS
0.1%
Published
Jun 2, 2026
CISA due date
—
The record
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
Affected packages
| Ecosystem | Package | Introduced | Fixed |
|---|
Related vulnerabilities
- CVE-2026-68554 Coturn is a free open source implementation of TURN and STUN Server. CWE-924
- CVE-2026-14681 Improper enforcement of message integrity in PostgreSQL GSSAPI support allows a user to… CWE-924
- CVE-2026-54891 Improper Enforcement of Message Integrity During Transmission in a Communication Channel… CWE-924
- CVE-2026-39827 An authenticated SSH client that repeatedly opened channels which were rejected by the… CWE-924
- CVE-2024-3596 RADIUS Protocol under RFC 2865 is susceptible to forgery attacks by a local attacker who… CWE-924